CTD-N Technical Manual
page 120
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
OTIME
Displays and sets the on time
Usage
OTIME <CR> or <LF>
Operating Modes
Open
Description
The OTIME command displays and allows you to set the on time of the instrument. The on time is
the time in hours, minutes and seconds beginning at the start of each interval time during which the
instrument is on and collecting data.
NOTE.
The time set by the OTIME command must be shorter than the time set by the ITIME
command by at least five seconds.
When setting the on time, use the hh:mm:ss format:
hh:
Hours—01 to 24
mm:
Minutes—00 to 59
ss:
Seconds—00 to 59
Returns
If no on time has been set:
NO ON time Enter time (hh:mm:ss):
If an on time has been set:
ON Time = 00:00:20 Enter time (hh:mm:ss):
Examples
Enter OTIME <CR> or <LF>
The instrument displays the current on time and allows you to enter a new time:
ON Time = 00:00:20 Enter time [hh:mm:ss]:
Enter a <CR><LF> to leave the on time as is, or enter the time in hours, minutes and seconds, and
then enter a <CR><LF>.
See
Also
ITIME, TIME, DATE