CTD-N Technical Manual
P/N 95F-6000-00 (February 2017)
page 121
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
RCAL
Displays the calibration constants
Usage
RCAL <CR> or <LF>
Operating Modes
Open
Description
The RCAL command displays all of the instrument calibration constants.
NOTE.
The ***E command must be entered to save any changed calibration constants. In
addition, the instrument serial number and the version number of the firmware cannot be
changed.
Returns
See Examples below.
Examples
Enter RCAL <CR> or <LF> The instrument displays the calibration constants:
S/N=1001 Firmware Version 2.2 CDATE=Not Set A1=0.0 B1=9.000000E-05 C1=0.0 D1=0.0
KFAC=1.360000 A2=0.0 B2=5.000000E-01 C2=0.0 D2=0.0 E2=0.0 F2=0.0
G2=0.0 H2=0.0 I2=0.0 A03=0.0 B03=0.0 C03=0.0 A503=1034.000000 B503=0.0 C503=0.0
A1003=2068.000000 B1003=0.0 C1003=0.0 A3=0.0 PS0=1.000000E-01 PS50=1034.400024
PS100=2068.800049
See Also
***C, ***E