20
CHAPTER FOUR: VEINVIEWER FLEX SETUP
020-300020-03 Rev. 1 (03-2013)
VEINVIEWER FLEX USER GUIDE
SUPER CLAMP: The Super Clamp is provided to allow mounting to flat surface
as well as alternate round or square surfaces such as a bed rail or IV pole.
WARNING: The health-care professional shall be responsible
for ensuring VeinViewer Flex and mobile system (such as an
IV Pole) is stable, and does not tip during set-up, operation,
transport, and storage, by making sure VeinViewer Flex and
S-Mount is properly mounted to IV Stands or IV Poles with
the specifications outlined on this page.
IV Stand Specifications
IV Pole Specifications
Max arm overhang: 28 in. (71.1 cm)
Max arm overhang: 28 in. (71.1 cm)
Max floor-to-Clamp height: 57 in.
(144.7 cm)
Max floor-to-Clamp height: 62 in.
(157.5 cm)
Min base weight: 16.5 lbs (7.5 kg)
Min base weight: 4 lbs (1.8 kg)
Min base size: 19.0 in. (48.3 cm)
Min base size: 23.25 in. (59.0 cm)
Pole diameter: 0.75 in. (1.9 cm) - 1.0
in. (2.5 cm)
Pole diameter: 0.75 in. (1.9 cm) - 1.0
in. (2.5 cm)
Min wheel diameter: 2.75 in. (7.0 cm) Min wheel diameter: 2 in. (5.0 cm)
Min number of wheels: 5
Min number of wheels: 5
WARNING: DO NOT leave VeinViewer Flex or S-Mount
PERMANENTLY mounted to Bed Rails. This condition could
result in bed entrapment.
V Block
Adapter
Clamp Configuration for
Mounting to a Flat Surface
Clamp Configuration for
Mounting to a Pole (Example;
IV Stand, IV Pole, Bed Rail)
Summary of Contents for VeinViewer Flex
Page 1: ...020 300020 03 Rev 1 02 2013 USER GUIDE VeinViewer Flex ...
Page 2: ......
Page 71: ...NOTES 65 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 NOTES ...
Page 72: ...66 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 73: ...NOTES 67 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 ...
Page 74: ...68 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 75: ......