CHAPTER TEN: TROUBLESHOOTING
45
VEINVIEWER FLEX USER GUIDE
020-300020-03 Rev. 1 (03-2013)
PROBLEM
STATUS INDICATOR
CAUSE
SOLUTION
VeinViewer Flex does not
power ON
Status Indicator is Off
VeinViewer Flex is turned OFF
Press the Power Button for a full second.
Battery does not have sufficient
charge
Replace with spare fully charged Battery, recharge primary Battery, or operate
VeinViewer Flex by plugging the AC Power Adapter into a standard power
receptacle.
Note: If the SUSPEND or ERROR Status Indicator on the Battery Charger is lit,
the battery is not charging properly. Remove Battery from Battery Charger per
"Battery Charging" procedure in CHAPTER FOUR: VEINVIEWER FLEX SETUP. If
the Battery still does not charge properly, contact your local distributor or
Christie Technical Support for replacement Battery.
Batteries are not making good
electrical contact
Be sure the Battery is properly secured inside the Battery compartment.
Wipe the Battery grooves with a CLEAN dry cloth.
Be sure the Battery Door is closed securely.
(If operating by AC power)
AC Power Adapter is damaged
DO NOT use or attempt to repair AC Power Adapter.
Immediately contact Christie Technical Support.
Summary of Contents for VeinViewer Flex
Page 1: ...020 300020 03 Rev 1 02 2013 USER GUIDE VeinViewer Flex ...
Page 2: ......
Page 71: ...NOTES 65 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 NOTES ...
Page 72: ...66 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 73: ...NOTES 67 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 ...
Page 74: ...68 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 75: ......