CHAPTER TEN: TROUBLESHOOTING
53
VEINVIEWER FLEX USER GUIDE
020-300020-03 Rev. 1 (03-2013)
Error pop-up while using
VeinViewer Flex Connect
None
Windows error
Follow on screen Windows instructions.
Disconnect VeinViewer Flex from the PC.
Restart the PC.
Reset VeinViewer Flex by removing the Battery and AC Power Adapter from
VeinViewer Flex.
Wait for 5 seconds, then reinstall the Battery and then Plug the AC Power
Adapter to VeinViewer Flex. Power up VeinViewer Flex.
Reconnect VeinViewer Flex to PC via USB Connection.
Restart the VeinViewer Flex Connect software.
If problem persists, document the error description and contact Christie
Technical Support.
PROBLEM
STATUS INDICATOR
CAUSE
SOLUTION