CHAPTER TEN: TROUBLESHOOTING
47
VEINVIEWER FLEX USER GUIDE
020-300020-03 Rev. 1 (03-2013)
No visible change in the
image field or
functionality when a
feature button
is pressed
Imaging mode icon in
image border
Control Panel button not
properly pressed
Firmly press and release alternate Control Panel buttons to verify Control Panel
functionality.
Note: Make sure, desired modes and features have been enabled using the
VeinViewer Flex Connect software.
See the "IMAGING MODES" section in
CHAPTER SIX: OPERATING INSTRUCTIONS for licensed imaging modes and
features.
Projected image appears
out of alignment
Drop Event Indicator
in image border and
Status Indicator
illuminated solid
orange. See CHAPTER
SIX: OPERATING
INSTRUCTIONS
VeinViewer Flex was accidently
dropped
See "DROP EVENT" in this chapter.
Projected image is out of
focus
Image border text
appears blurry
VeinViewer Flex is not set to
optimal focusing distance
See "FOCUSING PROCEDURE" in CHAPTER SIX: OPERATING
INSTRUCTIONS.
Projected image does not
properly display veins
None
VeinViewer Flex is being
operated outside or in direct
sunlight
Do not use the VeinViewer Flex outside or in direct sunlight. See the
recommended operating conditions in CHAPTER ELEVEN: TECHNICAL
SPECIFICATIONS.
PROBLEM
STATUS INDICATOR
CAUSE
SOLUTION
Summary of Contents for VeinViewer Flex
Page 1: ...020 300020 03 Rev 1 02 2013 USER GUIDE VeinViewer Flex ...
Page 2: ......
Page 71: ...NOTES 65 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 NOTES ...
Page 72: ...66 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 73: ...NOTES 67 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 ...
Page 74: ...68 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 75: ......