Conducted RF
IEC 61000-4-6
3 Vrms from 150 kHz to
80 MHz
3Vrms
Portable and mobile RF communications equipment should be used no
closer to any part of VeinViewer Flex, including cables, than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance: d=1.2 x √P from 150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m from 80 MHz to
2.5 GHz
3 V/m
d=1.2 x √P from 80 MHz to 800 MHz
d=2.3 x √P from 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance level in
each frequency range
b
.
Interference may occur in the vicinity of equipment marked with this
symbol.
Guidance and manufacturer’s declaration – electromagnetic immunity
(CONTINUED)
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
Summary of Contents for VeinViewer Flex
Page 1: ...020 300020 03 Rev 1 02 2013 USER GUIDE VeinViewer Flex ...
Page 2: ......
Page 71: ...NOTES 65 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 NOTES ...
Page 72: ...66 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 73: ...NOTES 67 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 ...
Page 74: ...68 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 75: ......