58
CHAPTER ELEVEN: TECHNICAL SPECIFICATIONS
020-300020-03 Rev. 1 (03-2013)
VEINVIEWER FLEX USER GUIDE
Guidance and manufacturer’s declaration – electromagnetic immunity
VeinViewer Flex is intended for use in the electromagnetic environment specified below. The customer or the user of VeinViewer Flex should ensure
that it is used in such an environment.
Immunity Test
Test Level
IEC 60601-1-2
Compliance Level
Electromagnetic Environment – Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
The floor should be wood, concrete, or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV Power Supply Lines
±1 kV input/output lines
±2 kV Power Supply Lines
Not Applicable
MAINS power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
±1 kV differential mode
Not Applicable
MAINS power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short interruptions,
and voltage variations on power
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(>60% dip in U
T
)
for 5 cycles
70% U
T
(>30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5s
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(>60% dip in U
T
)
for 5 cycles
70% U
T
(>30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5s
MAINS power quality should be that of a typical commercial or hospital
environment. If the user of VeinViewer Flex requires continued operation
during power MAINS interruptions, it is recommended that the
VeinViewer Flex be powered from an uninterruptible power supply (UPS)
or a battery.
Summary of Contents for VeinViewer Flex
Page 1: ...020 300020 03 Rev 1 02 2013 USER GUIDE VeinViewer Flex ...
Page 2: ......
Page 71: ...NOTES 65 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 NOTES ...
Page 72: ...66 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 73: ...NOTES 67 VEINVIEWER FLEX USER GUIDE 020 300020 03 Rev 1 03 2013 ...
Page 74: ...68 NOTES 020 300020 03 Rev 1 03 2013 VEINVIEWER FLEX USER GUIDE ...
Page 75: ......