Chapter 15 Troubleshooting and correcting faults
15.3 Emergency pulse mode
WARNING
If the Ikus is operating in emergency pulse mode, the user must
immediately visually check the blood pump(s) to determine whether the
pump(s) are filling and ejecting completely. If one pump is not filling
and/or ejecting completely the patient must be supported immediately
with the replacement
Ikus
. Use the manual pump while securing the
replacement Ikus (see section 15.4: Connecting the patient to a
replacement Ikus, page 202 and section 15.5: Driving blood pump(s)
with the manual pump, page 204). Otherwise there is the risk that the
patient will not be supported sufficiently.
If the emergency pulse mode is activated while the backup system is
already active, the
Ikus
is no longer able to drive both pumps. In this
case the patient must be supported immediately with the replacement
Ikus
. Use the manual pump while securing the replacement
Ikus
(see
section 15.4: Connecting the patient to a replacement Ikus, page 202
and section 15.5: Driving blood pump(s) with the manual pump,
page 204). Otherwise there is the risk that the patient will not be
supported sufficiently.
IMPORTANT: In emergency pulse mode a controlled shut down is not possible.
IMPORTANT: In emergency pulse mode it is not possible to acknowledge the acoustic
alarm.
If the graphs in the monitor program are not moving and the parameters cannot be
adjusted even after the laptop and the monitor program have been restarted, then the
Ikus
is operating in emergency pulse mode. Both control computers have failed and
are no longer communicating with each other. The emergency pulse board has taken
over control of the left and the right pneumatic system.
Replace the Ikus with a backup
Ikus
if possible. If no replacement
Ikus
is available, the
Ikus
will continue to support the patient in emergency pulse mode until a replacement
Ikus
is ready.
In emergency pulse mode the system operates with the following settings:
Synchronous mode (biventricular)
Systol.
pressure
[mmHg]
Diastol.
pressure
[mmHg]
Rate
[bpm]
relat. systole
duration [%]
left
210
-40
70
40
right
150
-40
70
40
Tab. 15-2
Settings in emergency pulse mode
Always immediately …
HOTLINE
Notify Berlin Heart! 866.249.0128
1000721x09 Revision 8
200
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...