
Chapter 6 Instructions for use: Ikus
To transport the
Ikus
, only the supplied original transport crate must be
used.
Make sure that the
Ikus
is standing firmly and securely in the crate and
that the crate is closed and sealed properly.
The crate must always be transported as marked. Do not tip the
Ikus
driving unit stored in the transport crate or turn it upside down.
Always observe a resting period of 6 hours after each transportation
before switching on the
Ikus
!
NOTICE
During storage and transportation, the power switch (toggle switch)
should be in
[I]
position.
To transport the
Ikus
, only use the original transport crate supplied.
Ensure that the
Ikus
is placed firmly and securely inside the transport
crate.
Ikus Accessories
• 1 Instruction for use (supplied separately)
• 2 Tank units (supplied separately)
•
2 Keys
•
1 Power cable
• 1 Alarm connector
• 1 Envelope with password
Also see section 16.4: Sample copy: Ikus Incoming Checklist, page 221.
6.7.1 Unloading the Ikus from the shipping crate
NOTICE
Do not discard the shipping crate!
INSTRUCTION
1.
Follow inspection steps per
Ikus Incoming Checklist
.
2.
Place the crate with its back against a wall and/ or lock the brakes to prevent
the crate from moving during unloading.
3.
Open the door latches. Flip open the secure latch and turn counter clockwise
to disengage latch from front door.
4.
Swing front door open to the left.
5.
Unlatch the ramp safety retainer.
6.
Open the ramp while holding the black securing plate behind the ramp.
7.
Open the front securing plate (is connected to the crate with a rope) and place
the plate to the right of the crate.
1000721x09 Revision 8
85
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...