Chapter 14 Error Messages and corrective measures
Restart the Ikus after the start-up test if one of the above messages appears
INSTRUCTION
1.
According to display: If necessary: enter
<7>
to read any messages; go back
by pressing
<Enter>
. Important: enter
<1>
or
<x>
to exit the monitor program.
A 2. window appears.
2.
2. window: According to the display, select the option
End
. The monitor pro
gram writes a log file and is terminated. Wait until the message
Switch off
drive with main switch!
appears.
3.
Switch off the main switch (key switch) and then switch it back on again.
If the Ikus now ends the start-up test without a message:
Start up the Ikus.
If one of the above messages appears again:
INSTRUCTION
1.
Do not start up the Ikus. Provide the patient immediately with a replacement
Ikus.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.26 Discrepancy in pressure measurement: system 1 (or
system 2 or 3)
•
Discrepancy in pressure measurement: system 1 (left)
•
Discrepancy in pressure measurement: system 1 (right)
•
Discrepancy in pressure measurement: system 1 (backup)
This message is only an information message.
INSTRUCTION
1.
Check the status of the patient.
2.
Check the filling and emptying of the blood pump(s), and monitor the function
of the Ikus. Do not operate the Ikus without supervision!
3.
Compare nominal values with the values actually generated. In the event of
discrepancies from high set values, if necessary reduce rate accordingly and/
or vary relative systolic duration, lower systolic driving pressure (e.g. from 210
mmHg to 200 mmHg), and/or raise the diastolic driving pressure parameters
(e.g. from -40 mmHg to -45 mmHg) if appropriate.
4.
Scroll down message window to check if message has previously appeared.
5.
Continue to observe message window to see if message reappears.
The information message
Discrepancy in pressure measurement: system 1 (2 or
3)
appears when the compressed air being produced at a constant rate by the
compressor in the present pump cycle is unable to meet the compression levels
required to achieve the defined parameters.
1000721x09 Revision 8
188
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...