4
Chapter 4 Description: blood pump, cannulae and accessories
Description: blood pump, cannulae and
accessories
EXCOR is an extracorporeal electro-pneumatically driven ventricular assist device. It
can be used for either univentricular or biventricular support. EXCOR is comprised of
the following permanently active components:
- extracorporeal blood pump(s)
- inflow and outflow cannula(e)
- 1 driving tube for each blood pump
- Ikus
1
1
cannula
2
blood pump
2
3
driving tube
4
Ikus
3
4
Fig. 4-1
EXCOR shown in situ as a biventricular assist device in pediatric application
Overview
The blood flows from the atrium or the ventricle through the inflow cannula into the
blood chamber of the pump and then from this blood chamber through the outflow
cannula into the aorta or into the pulmonary artery. A driving tube is used to connect
the air chamber of the pump to the electro-pneumatic Stationary Driving Unit Ikus. Ikus
generates the suction and driving pressures required to move the triple-layer
membrane separating the blood chamber from the air chamber.
1000721x09 Revision 8
53
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...