Chapter 11 Wound care and treatment
Checking the pump areas which come in contact with blood
1
transition inflow cannula – inflow connector
2
inflow stub in front of inflow valve
3
inflow valve
4
inflow stub behind inflow valve
5
area between inflow and outflow stubs
6
remaining area of blood chamber
7
transition blood chamber - membrane (directly
above the reinforcement ring)
8
outflow stub in front of outflow valve
9
transition outflow connector – outflow cannula
10
outflow valve
11
outflow stub behind outflow valve
Fig. 11-18
Diagram of EXCOR blood pump (top view of blood chamber)
ADVICE
During the visual check, first clean the blood pump then illuminate the
blood chamber with a flashlight. This makes it easier to detect deposits.
Enter all of the findings into the blood pump log. (see section 16.6:
Sample copy: EXCOR pump log, page 227).
Cautionary measures
Initial signs of deposits: check anticoagulation therapy and adjust therapy if
necessary.
Floating deposits inside the pump: replace the pump!
11.5.3 Checks using the monitor program
Record all drive parameters and adjust if necessary.
Objective: the blood pump(s) must fill and eject completely in each pumping cycle, the
diastolic pressure should be as low as possible.
ADVICE
Record the parameter values once a day.
To record the parameters use the sample copy in section 16.7: Pump
performance flow sheet, page 230.
1000721x09 Revision 8
143
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...