Chapter 7 Commissioning the Ikus and setting parameters
INSTRUCTION
1.
Place the cursor with
<
↑
>
in the parameter table and navigate with
<
←
>
/
<
→
>
to the desired field in the parameter table. Adjust with
<
↓
>
,
<
↑
>
/
<Bild-
↓
>
,
<
Bild-
↑
>
then confirm with
<Enter>
.
2.
Set the start-up parameters:
Systole [mmHg]
left/ right
Diastole
[mmHg]
left/ right
Rate [bpm]
Rel. systol.
duration [%]
left/ right
100/ 80
-5/-5
30
40/40
Tab. 7-2
Start-up parameters
7.2.7 Connecting the blood pump(s) to the Ikus
WARNING
Do not kink either the driving tubes or the cannulae.
NOTICE
State of the blood pumps when they are initially connected: filled with
sterile injectable saline, de-airing needle in place. To allow easier
handling, the driving tubes are not connected until the inflow and
outflow cannulae have been connected to the pump (see section 9.9:
Connecting the blood pumps to the cannulae, page 128).
INSTRUCTION
1.
Open the driving tube connector marked in red (univentricular) or both connec
tors (biventricular). To do so, pull the seal plugs out of the connector(s).
2.
Connect the driving tube to the Ikus. To do so, push the plug of the driving tube
into the connector. The sound of the plug snapping into place is clearly audi
ble. Check that the plug is securely connected. To do so, grip the plug body
above the release sleeve and pull on it. Do not pull from the release sleeve,
and never from the tube!
3.
In biventricular mode: observe the color of the markings.
4.
In biventricular mode: repeat the procedure for the second pump.
Operating mode
Ikus connector
biventricular
LVAD: connector marked red
RVAD: connector marked blue
univentricular
connector marked red
Tab. 7-3
Assignment: operating mode, blood pump, connector
1000721x09 Revision 8
101
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...