Chapter 14 Error Messages and corrective measures
The message can have the following causes:
•
Extreme parameter values have been set (extremely high values, extremely low
values)
•
Restricted pneumatic performance (maintenance required)
•
System malfunction
If the message appears several times within 24 hrs
NOTICE
Damage to the Ikus is technically impossible when this message
appears!
Avoid a further increase in the rate or, alternatively, the driving pressure
for parameters set at high values. Otherwise the Ikus might perform a
test phase, which means that the blood pump of the corresponding
system is briefly stopped for the duration of this test
The information message Discrepancy in pressure measurement:
system 1 (2 or 3) is not accompanied by an acoustic signal.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.27 Communication with Laptop failed
This message informs the user that the communication between the control
computers and the monitor program has been temporarily interrupted. This occurs if
the monitor program is exited, for example to administrate user IDs and passwords.
Since this message only appears when the communication is re-established, this
message does not require user intervention.
1000721x09 Revision 8
189
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...