Chapter 6 Instructions for use: Ikus
Fig. 6-19
Pump size and single-step mode
3.
The cursor is located on the
OK
field. In order to restart the driving unit with
the current settings, press the
<Enter>
key. IMPORTANT: If the user leaves
the
OK
field before pressing
<Enter>
, at least once
Step left
or
Step right
,
respectively has to be performed again. Only now the
OK
field is highlighted
again and it is possible to confirm
OK
by pressing
<Enter>
Select the pump size, page 98).
6.3.2 Pause left/ Pause right: stopping an individual blood pump
IMPORTANT: The option
Pause right
is only available in biventricular mode. The
option is displayed in univentricular mode, but it is not activated. It is only possible to
select
Pause left
.
INSTRUCTION
1.
Select
Pause left
or
Pause right
, as required, then press
<Enter>
to confirm.
Respond to the prompt in the dialog window by pressing the
<X>
key or the
<1>
key. The selected pump will stop. The view
Pump size and single-step
mode
is displayed.
2.
To restart the pump, at least once
Step left
or
Step right
, respectively has to
be performed (see section 7.2.3: Select the pump size, page 98).
6.3.3 Drive OFF: switching the Ikus off
WARNING
The Ikus power switch (toggle switch) must always be in the
[I]
position
even if the Ikus is set to
[0]
position at the main switch (key switch)!
CAUTION
The sequence of operations below must always be followed.
Do not switch the Ikus off unless the batteries are fully charged (i.e. all
yellow charge indicator LEDs are on).
Keep all driving tube connectors covered at all times when not in use.
1000721x09 Revision 8
78
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...