Chapter 17 EMC tables
Immunity test IEC 60601
test level
Compliance level Electromagnetic
environment -guidance
Electrostatic
± 6 kV contact
± 6 kV contact
Floors should be wood,
discharge
(ESD)
± 8 kV air
± 8 kV air
concrete or ceramic tile.
If floors are covered with
IEC 61000-4-2
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
< 5 % UT
< 5 % UT
Mains power quality
short
(>95 % dip in UT) (>95 % dip in UT) should be that of a
interruptions
for 0.5 cycle
for 0.5 cycle
typical commercial or
and voltage
variations on
power supply
input lines
40 % UT
(60 % dip in UT)
for 5 cycles
40 % UT
(60 % dip in UT)
for 5 cycles
hospital environment. If
the user of the Ikus
requires continued
operation during power
IEC 61000-4-11
70 % UT
(30 % dip in UT)
70 % UT
(30 % dip in UT)
mains interruptions, it is
recommended that the
for 25 cycles
for 25 cycles
Ikus be powered from an
< 5 % UT
< 5 % UT
uninterruptible power
(>95 % dip in UT) (>95 % dip in UT) supply or a battery.
for 5 sec
for 5 sec
Power
3 A/m
100 A/m
The Ikus can be used up
frequency (50/
to 1 m from power
60 Hz)
cables carrying up to
magnetic field
100 A.
IEC 61000-4-8
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Tab. 17-2
Electromagnetic immunity - part 1
17.4 Electromagnetic immunity - part 2
The Ikus is intended for use in the electromagnetic environment specified below. The
customer or the user of the Ikus should ensure that it is used in such an environment
1000721x09 Revision 8
232
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...