Chapter 9 Implantation - surgical procedure
9.7.2
Anastomosis of the arterial cannula
Aorta
INSTRUCTION
1.
Tangentially clamp the ascending
aorta and make a longitudinal
opening of a length which is suit
able for the cannula diameter. If
necessary, offset the incision later
ally to the right by up to 45°.
2.
Anastomose the cannula using ten
teflon-backed double-reinforced
individual monofilament (e. g. 4-0
EB) U-sutures. (If simpler condi
tions are encountered, a running
suture can be made instead.)
3.
Remove all air from the cannula
and use a tube clamp to clamp the
cannula below the anastomotic site.
If it is necessary to clamp any other
part of the cannula that is not cov
ered with velour, cover the part of
the cannula that will be clamped
with a gauze sponge.
Pulmonary artery
Fig. 9-11
Anastomosis of the aortic cannula
INSTRUCTION
1.
Make a longitudinal incision of a
size suitable for the cannula diame
ter in the pulmonary artery.
2.
Anastomose the cannula using 10
teflon-backed, double-reinforced
individual monofilament (e. g. 4-0
EB) U-sutures. (If simpler condi
tions are encountered, a running
suture can be made instead.)
3.
Remove all air from the cannula
and use a tube clamp to close it
below the anastomosis. If it is nec
essary to clamp any other part of
the cannula that is not covered with
velour, cover the part of the cannula
that will be clamped with a gauze
sponge.
Fig. 9-12
Cannulation of the pulmonary artery
1000721x09 Revision 8
127
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...