Chapter 14 Error Messages and corrective measures
3.
Provide the patient immediately with a replacement Ikus and switch off the
malfunctioning Ikus. (see section 15.4: Connecting the patient to a replace
ment Ikus, page 202). If no replacement Ikus is available: Support the patient,
if necessary (in the event of malfunction of the defective Ikus), by means of the
manual pump! (see section 15.5: Driving blood pump(s) with the manual
pump, page 204)
HOTLINE
Notify Berlin Heart! 866.249.0128
14.23 Problem: <<Text>>
If this message appears during the start-up test::
see section 14.25: Error messages during the start-up test, page 187.
If this message appears outside of the start-up test,
a serious problem exists.
The Ikus has detected a serious problem.
INSTRUCTION
1.
Check the status of the patient.
2.
Check the filling and emptying of the blood pump(s), and monitor the function
of the Ikus. Do not operate the Ikus without supervision!
3.
Provide the patient immediately with a replacement Ikus and switch off the
malfunctioning Ikus. (see section 15.4: Connecting the patient to a replace
ment Ikus, page 202). If no replacement Ikus is available: Support the patient,
if necessary (in the event of malfunction of the defective Ikus), by means of the
manual pump! (see section 15.5: Driving blood pump(s) with the manual
pump, page 204)
HOTLINE
Notify Berlin Heart! 866.249.0128
14.24 Self-test is not completed by passive computer!
The passive processor was unable to end the self-test of the alarm circuit.
If, within 8 seconds, the message Alarm circuit test OK appears,
the Ikus is working perfectly. No further measures are necessary.
If the message Alarm circuit test OK does not appear within 8 seconds,
the alarm circuit is defective. A possible error might not have been detected.
The Ikus has detected a serious problem.
INSTRUCTION
1.
Check the status of the patient.
2.
Check the filling and emptying of the blood pump(s), and monitor the function
of the Ikus. Do not operate the Ikus without supervision!
1000721x09 Revision 8
186
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...