Chapter 15 Troubleshooting and correcting faults
7.
Respond to the prompt in the dialog window by pressing the
<X>
key or the
<1>
key. The system stops operation immediately and writes an operating log.
8.
Wait until the log has been completed (message:
Switch off drive with main
switch!
). IMPORTANT: Do not switch the
Ikus
off yet.
9.
Shut down the monitor program. Press
<F10>
and confirm by pressing the
<X>
key or the
<1>
key. The start menu is displayed on the laptop.
10.
Insert the USB stick in the laptop before the laptop is switched on. In the start
menu, select
4. Save data
(
<4>
). The LOG files are copied onto the USB stick.
After that the start menu appears again.
11.
Switch off laptop. Take the USB stick out of the port (but never when the laptop
is switched on).
12.
Send the LOG files by e-mail to [email protected].
15.4 Connecting the patient to a replacement Ikus
CAUTION
The
Ikus
must always be connected to the power supply when it is
switched on. This is the only way to ensure that the start-up test is
performed completely and possible malfunctions can be detected.
This is necessary if …
• maintenance is required
•
the
Ikus
is defective
Ikus and all of its parts shall not be serviced or maintained while in use with a patient.
Switching the replacement Ikus on
INSTRUCTION
1.
Prepare the replacement
Ikus
and connect it to the mains. Secure the mains
cable with the plug clip. Ensure that the power switch (toggle switch) is set to
[I]
position.
2.
Use the seal plugs to seal both driving tube connectors.
3.
Switch on the replacement
Ikus
. To do so, turn the key switch to the
[I]
posi
tion. The battery charge indicator will light up and the number of hours the driv
ing unit has been operated to date will be displayed. The mains operation
indicator lights up.
4.
Switch the laptop on. The menu
Select language
appears.
Setting the parameter values of the replacement Ikus
NOTICE
If no parameter values are entered into the replacement
Ikus
, the
replacement
Ikus
starts up with the following default parameter values
(standard parameters):
1000721x09 Revision 8
202
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...