Chapter 11 Wound care and treatment
11.5.4 Replacing the blood pump due to growth of the patient
CAUTION
In children, plan to replace the pump(s) with a larger pump(s) in good
time, to prevent the possibility of inadequate support due to an
insufficient discharge rate.
The pump selected at the time of transplantation may not be adequate for the entire
period of cardiac support. Growth and/or weight gain can result in the patient not
receiving adequate support. Use the chart in section 16.1.2: Overview: Relationship:
body weight – pump size, page 211, to plan, in good time, which pump(s) the patient
may need to change over to. This chart is for guideline purposes only and is not
binding for each individual case. This decision must be taken by the surgeon in
consultation with Berlin Heart GmbH.
HOTLINE
Notify Berlin Heart! 866.249.0128
The blood pump(s) must be replaced as described in section 15.1: Replacing the
blood pump(s), page 195.
1000721x09 Revision 8
144
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...