Chapter 9 Implantation - surgical procedure
If it is necessary to apply a clamp directly to the cannula in order to pull
the cannula through the skin, the following procedures should be
observed:
- Position the clamp at the distal end of the cannula
- After the cannula has been pulled through the skin, cut off and
discard the part of the cannula where the clamp was applied.
If it is necessary to clamp any other part of the cannula (cannula
extension set / connecting set resp.) that is not covered with velour,
cover the part that will be clamped with a gauze sponge.
If replacement of an EXCOR blood pump is required, the following
procedures should be observed:
- The cable tie covering the EXCOR cannula on the stub of the blood
pump should be removed carefully. Use an appropriate blunt tool.
IMPORTANT: never use a sharp instrument, for example, a scalpel or
scissors, to remove the cable tie. This may cause damage to the
cannula.
- If a cannula extension set/ connecting set needs to be cut for a
pump replacement, ensure that there will be sufficient length of the tube
part remaining to meet the minimum length recommendations. See
Tab. 9-1, page 120.
Do not kink the drivelines. Otherwise there might not be sufficient pump
output
Do not kink the cannulae needlessly. Otherwise there might not be
sufficient pump output. Moreover, cannulae might be damaged.
At least daily, the EXCOR cannulae should be inspected for signs of
wear or damage. ADVICE: To avoid needless kinking of the cannulae
use a mirror for inspection of the bottom side of the blood pump.
In no case should the cannulae either be kinked directly at the
connector to the blood pump or at the transition area between velour
and silicone.
9.3.1 Description: Cannula extension and connecting sets
Berlin Heart supplies connecting sets to bridge different connector diameters in the
blood pump and cannula (6/9, 9/12). This allows for greater flexibility when combining
blood pumps and cannulae. The connecting set may be used during implantation or
during the further course of therapy.
Berlin Heart supplies cannula extension sets for blood pumps/cannula combinations
with diameters of 6/6, 9/9 and 12/12.
The cannula extension set is used to lengthen the piece of cannula which remains
after the cannula has been shortened.
1000721x09 Revision 8
118
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...