Chapter 13 Weaning and Explantation for BTR and BTT
10 ml / 15 ml pump Weaning Sequence
Day 5
After confirmation of eligibility criteria, the following steps should be
performed in the cath lab under echo guidance
1
:
1. Obtain standard access for RHC (if possible with out sedation).
2. Administer UFH 75 units/kg x _____ kg = _____ mg IV x 1 [max 5000
units].
3. After 5 minutes, reduce the pump rate step-wise from WR (50 bpm)
to 30 bpm in increments of 5 bpm q5 min. After 5 minutes at 30 bpm,
reassess LV size and function.
4. After a total time of
10 min
at 30 bpm, stop the pump for
3 min
and
reassess LV size and function. During pump stoppage, use the
manual pump to pump twice q30 seconds to minimize the risk of
thrombus formation due to blood stagnation while Ikus is
disconnected.
5. After 3 minutes, initiate norepinephrine infusion at 0.01 mcg/kg/min IV
gtt titrated to MAP 20% above baseline x 5 min. While doing so,
proceed pumping manually twice q30 seconds.
6. If LV size and function acceptable, proceed pumping manually twice
q30 seconds for
3 min
. While doing so, reassess LV size & function,
and record RAP, PAP, PCWP and MVS.
7. After 6-minute pump stop: If the patient meets all eligibility criteria,
reconnect pump to Ikus and resume pumping at 50 bpm until the
actual surgical procedure of explantation takes place. If the patient
does not meet all criteria, reconnect Ikus and resume pumping at IR.
Did the patient satisfy all 5 eligibility criteria
throughout the period of weaning and pump
stoppage?
ÿ
NO -STOP
ÿ
YES
Proceed
MD________
Tab. 13-3
10 ml / 15 ml pump weaning sequence
1
TEE unless echo windows insufficient. The last weaning increment may be less
than 5 bpm if the wean interval is not a multiple of 5.
13.1.6 25 / 30 ml pump
The individual weaning progress is based upon the following parameters:
Parameter
Explanation
Abbr.
Value
initial rate
rate prior to any
weaning
IR
Please enter:
weaning rate
lowest rate achieved
during weaning
process, depends
on pump size
WR
40 bpm
Tab. 13-4
Important parameters for weaning progress
1000721x09 Revision 8
158
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...