Chapter 9 Implantation - surgical procedure
During implantation/ When replacing a blood pump
INSTRUCTION
1.
Connect the cannula extensions (the connectors resp.) with the blood pump.
To do so, push the section of the cannula extension (the connector resp.) onto
the titanium connector of the blood pump. Prime the blood pump with the can
nula extensions (with connectors resp.).
2.
Connect the cannula extensions (the connectors resp.) with the cannulae. To
do so, push each cannula onto the titanium connectors of the cannula exten
sion (of the connectors resp.) while flushing with sterile injectable saline solu
tion.
3.
Proceed according to context. See section 8.4: Moving the membrane to the
end-of-diastole position, page 112 and section 15.1: Replacing the blood
pump(s), page 195.
Without replacing a blood pump
INSTRUCTION
1.
Push the free end of the cannula onto the titanium connector of the cannula
extension (of the connector resp.).
2.
Flush the tube sections with sterile injectable saline solution.
3.
Push the sections of tube, which are free of air, onto the titanium connectors of
the blood pump.
4.
Proceed according to context. Act according to 15.1.2, page 196 and 15.1.3,
page 197 respectively but without replacing the blood pump.
Securing the connections
INSTRUCTION
1.
Secure each connection between silicone tube and titanium connector with at
least 1 cable tie. See section 9.11: Securing the connections, page 130.
9.4
Access
INSTRUCTION
1.
Median sternotomy. Make sure that there is absolutely no bleeding.
2.
Insert standard cardiopulmonary bypass cannulae (bicaval cannulation).
3.
Initiate extracorporeal circulation.
4.
Place a vent in the left atrium, if necessary.
9.5
LV apex cannula
Refer to section 9.2: Use of the cannula tunneling tip, page 117.
1000721x09 Revision 8
121
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...