Chapter 8 Implantation: Preparations in the operating room
8.4
Moving the membrane to the end-of-diastole position
1
1
de-airing nipple (blood chamber)
2
driving tube connector (air chamber)
2
Fig. 8-6
De-airing nipple and driving tube connector
INSTRUCTION
1.
Pick up adapter tube, disposable syringe (membrane set) and the pump.
2.
Connect the adapter tube to the disposable syringe.
3.
Connect the free end of the adapter tube to the driving tube connector of the
blood pump.
4.
Remove all air from the air chamber of the pump. The blood pump membrane
is now in the end-of-diastole position.
5.
Seal the adapter tube with a tube clamp in order to keep the membrane in this
position.
8.5
De-airing the blood pump
WARNING
Make sure that no particles or liquids enter the air chamber of the blood
pump. Otherwise, the membrane may be damaged and the patient may
not receive adequate support.
Prepare and place the following ready for use:
• blood pump(s) with pump seal(s)
• 1 de-airing set (trocar and a de-airing tube) for each blood pump
• 50 ml disposable syringe for each blood pump
1000721x09 Revision 8
112
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...