
Chapter 15 Troubleshooting and correcting faults
INSTRUCTION
1.
Bring membrane to the end-of-diastole position, position de-airing needle,
rinse and fill pump with sterile injectable saline (see section 8.4: Moving the
membrane to the end-of-diastole position, page 112 and section 8.5: De-airing
the blood pump, page 112).
2.
Connect the driving tube to the respective driving tube connector of the pump.
3.
Place the pump, ready for connection, with the titanium connectors pointing
upwards.
15.1.2 Replacing the right blood pump (RVAD/ BVAD)
Material
•
1 prepared replacement blood pump (see section 15.1.1: Preparing a
replacement blood pump, page 195)
•
1 tube connecting set (cable tie, cable-tie gun), included in the accessory set.
Only the cable ties and cable tie guns provided should be used.
Stopping the right blood pump and detaching the blood pump from Ikus
INSTRUCTION
1.
Bring the patient into the Trendelenburg position.
2.
The cable tie covering the EXCOR cannula on the stub of the blood pump
should be removed carefully. Use an appropriate blunt tool.
Important: never use a sharp instrument, for example, a scalpel or scissors, to
remove the cable tie. This may cause damage to the cannula. Check cannulae
immediately to make sure they are not damaged.
3.
If necessary log into the monitor program by entering user ID and password,
confirming the password with
<Enter>
.
4.
BVAD: Reduce rate of left blood pump to 30 bpm. Use
<
←
>
/
<
→
>
to navigate
cursor to the respective field of the parameter table, then use
<
↓
>
to adapt
value. Confirm with
<Enter>
.
5.
In the monitor program, select the option
Pause left
respectively
Pause right
and press
<Enter>
to confirm. Respond to the prompt in the dialog window by
pressing the
<X>
key or the
<1>
key. The right blood pump will stop.
RVAD:
Pause left
BVAD:
Pause right
The view
Pump size and single-step mode
is displayed.
6.
As soon as the right pump has stopped, clamp off the cannulae beneath the
right pump to be replaced and slide the cannulae off the pump. If it is neces
sary to clamp any other part of the cannula that is not covered with velour,
cover the part of the cannula that will be clamped with a gauze sponge.
7.
Check cannulae for visible deposits. If necessary, remove these deposits care
fully.
8.
Remove the driving tube of the pump to be replaced from the connector. To do
so, take hold of the release sleeve and pull this out of the connector.
1000721x09 Revision 8
196
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...