Chapter 4 Description: blood pump, cannulae and accessories
7
6
4.1
EXCOR blood pumps
1
1
driving tube connector
2
triple-layer membrane
3
blood chamber (de-airing nip
ple at back of pump)
8
9
4
arrow mark: indicates blood
2
flow direction
3
5
inflow stub
4
6
titanium connector: inflow stub
– inflow cannula
5
7
titanium connector: outflow
stub – outflow cannula
8
outflow stub
9
air chamber
Fig. 4-2
60 ml blood pump
EXCOR blood pumps have a transparent polyurethane (PU) housing which is divided
into an air chamber and a blood chamber by a triple-layer membrane.
The blood chamber has an inflow and an outflow stub to which the inflow and outflow
cannula, respectively, are connected. The pump stubs themselves are made of
polyurethane, the end of each stub is fitted with a titanium connector to which the
cannula will be connected. The valves located in the pump stubs keep the blood
flowing in one direction. EXCOR blood pumps are available with three-leaflet valves
made of polyurethane (10 - 60 ml stroke volume).
All surfaces of the pump coming into contact with the blood are coated with a
Carmeda
®
BioActive Surface (CBAS) coating. The transparent casing of the blood
pump allows easy visual monitoring of the filling and emptying of the blood chamber.
The blood pump is equipped with a de-airing nipple which is used for de-airing the
blood chamber when the pump is being commissioned.
The air chamber of the pump is equipped with a driving tube connector. This connector
is used to connect the blood pump to the driving tube through which air is pumped from
the Ikus. Ikus generates the suction and driving pressures required to move the blood
pump’s triple-layer membrane. A graphite powder layer is located between the
membrane layers in order to minimize friction.
4.2
EXCOR cannulae
3 different types of cannulae are available for EXCOR in various sizes for each type:
• atrial cannulae (as inflow cannulae)
• LV apex cannulae (as inflow cannulae)
• arterial cannulae (as outflow cannulae)
The cannulae are made of tissue-friendly silicone. Polyester-velour suture rings
enable convenient and safe anastomosis of the cannulae. The mid section of all
cannulae is covered with polyester-velour in order to promote good ingrowth of the
cannulae where they pass through the skin.
Some arterial cannulae have a shaping wire which allows the cannulae to be adapted
to each individual patient’s anatomic conditions.
1000721x09 Revision 8
54
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...