Chapter 13 Weaning and Explantation for BTR and BTT
25 / 30 ml pump Weaning Sequence
Day 3
After confirmation of eligibility criteria, the following steps should be
performed under echo guidance:
1
1.
Administer UFH 75 units/kg x _____ kg = _____ mg IV x 1 [max 5000
units].
2.
After 5 minutes, reduce the pump rate step-wise from RR2 (___ bpm)
to 30 bpm in increments of 5 bpm q 5 min. After 5 minutes at 30 bpm,
reassess LV size and function.
3.
Initiate exercise with gentle age-appropriate play tasks (e.g. patty
cake) as clinically appropriate, where possible
4.
After a total time of
30 min
at 30 bpm, stop the pump for
10 min
and
reassess LV size and function. During pump stoppage, use the
manual pump to pump twice q30 seconds to minimize the risk of
thrombus formation due to blood stagnation while Ikus is
disconnected.
5.
After 10-minute pump stop, reconnect pump to Ikus and resume
pumping at RR3 (___ bpm).
Did the patient satisfy all 5 eligibility criteria
throughout the period of weaning and pump
stoppage?
ÿ
NO - STOP
ÿ
YES -
Proceed
MD________
Da
y 4
After confirmation of eligibility criteria, the following steps should be
performed under echo guidance:
1
1.
Administer UFH 75 units/kg x _____ kg = _____ mg IV x 1 [max 5000
units].
2.
After 5 minutes, reduce pump rate step-wise from RR
3
(___ bpm) to
30 bpm in increments of 5 bpm q 5 min. After 5 minutes at 30 bpm,
reassess LV size and function.
3.
Initiate exercise with gentle age-appropriate play tasks (e.g. patty
cake) as clinically appropriate, where possible.
4.
After a total time of
30 min
at 30 bpm, stop the pump for
15 min
and
reassess LV size and function. During pump stoppage, use the
manual pump to pump twice q30 seconds to minimize the risk of
thrombus formation due to blood stagnation while Ikus is
disconnected.
5.
After a 15-minute pump stop: If the patient meets all eligibility criteria,
reconnect pump to Ikus and resume pumping at WR (40 bpm). If the
patient does not meet all criteria, reconnect Ikus and resume
pumping at IR.
Did the patient satisfy all 5 eligibility criteria
throughout the period of weaning and pump
stoppage?
ÿ
NO - STOP
ÿ
YES -
Proceed
MD________
Tab. 13-6
25 / 30 ml pump Weaning Sequence
1000721x09 Revision 8
161
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...