Chapter 14 Error Messages and corrective measures
If the message appears several times
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.13 Parameter set update failure
The Ikus cannot store the changed parameter values in the internal system memory.
The driving unit is operating with the changed values, but if a reset were to be
performed, the old values would be valid again.
INSTRUCTION
1.
Check the parameter values regularly. After a reset, if the Ikus continues to
work with the old parameter values: Re-adjust the parameter values.
If the message appears again:
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.14 Temperature sensors: <<8-digit binary code>>
•
Temperature sensors: <<8-digit binary code>>
WARNING
Do not use water or other liquids to cool the Ikus! Otherwise there is a
risk of short circuit and/or device malfunction.
IMPORTANT: An alarm was triggered by one of the sensors and an user muted it
(mute interval: 10 min). During this mute interval now another sensor generates
another alarm which is also muted by the user. This means that the remaining mute
time of the first triggered alarm will be extended for another full mute interval.
INSTRUCTION
1.
Determine whether internal or external influences have caused the driving unit
to over-heat. Is it exposed to direct heat from external sources? Is the ambient
temperature too high?
2.
If possible, remedy the situation (move the Ikus away from the heater, etc.).
Provide adequate ventilation. Acknowledge the message. The Ikus takes a
few minutes to cool down. Acknowledge the message repeatedly if necessary.
Usually, overheating of the Ikus is due to external factors such as direct thermal
radiation (e. g. direct sunlight or from heaters). Overheating may also be caused by
an internal fault, but this rarely occurs.
1000721x09 Revision 8
181
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...