Chapter 16 Appendix
Electro-pneumatic extracorporeal ventricular assist device
EXCOR
®
Pediatric VAD with Stationary Driving Unit Ikus
Input voltage
AC 115 V
Frequency
60 Hz
Power drawn
575 VA
Mains fuse
5 A
Mains cable
10 A, hospital grade
Connector External alarm
electrical data: max 1 A, 24 V
insulation specifications: 2.5 mm/ 4 mm
clearance and creepage distance between
alarm contact and 24 V extra low voltage
inside the device (coil-sided)
insulation test voltage: 500 V
Protection class
IPX1 (protection against touching live parts
not tested, tested safety from vertically
dripping water)
Pump rate
30 to 150 bpm
Sound level of acoustic alarm
71 dB (A)
Systolic pressure:
60 to 350 mmHg
Diastolic pressure
-100 to 0 mmHg
Pressure display accuracy
±
10%
Relative systolic duration
20 % to 70%
Off-mains operating time
max. 30 minutes
Battery charging time
6 h
Maintenance interval
2000 operating hours (at the latest after 6
months). In the event of permanently higher
ambient temperatures than recommended,
the maintenance intervals can shorten
drastically.
Product life Ikus
max. 8 years
Tab. 16-11
Technical specifications
1000721x09 Revision 8
219
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...