Chapter 7 Commissioning the Ikus and setting parameters
7.2.3 Select the pump size
NOTICE
It is not possible to select a larger blood pump and/or a higher rate for
the right pump than for the left one. The list of pump sizes available for
the right pump is limited accordingly.
The sizes written in red ink in the pop-up menu
Select pump size
can
be chosen. However, they are not the recommended option for each
individual case (see Fig. 7-4, page 98).
In individual cases, consideration is to be given as to whether a
combination that is not recommended is to be selected. The final
decision on the combination of blood pumps and cannulae is to be
reached by the implanting surgeon, in consultation with Berlin Heart,
Inc., Clinical Affairs.
Fig. 7-4
Select the pump size
INSTRUCTION
1.
In the pop-up menu, select the desired pump size with
<
↓
>
/
<
↑
>
.
2.
In biventricular mode: Using the cursor keys
<
←
>
/
<
→
>
, move to the field
marked right pump, select the desired pump size with the cursor keys
<
↓
>
/
<
↑
>
.
7.2.4 Select the cannula size
NOTICE
The sizes written in red ink in the pop-up menu
Select the inflow and
outflow cannula sizes
can be chosen. However, they are not the
recommended option for each individual case (see section 16.1.10:
Maximum rates for the pump-cannula combinations, page 216 and
section 16.1.11: Blood pump combinations in biventricular mode,
page 217). In individual cases, consideration is to be given as to
whether a deviant combination is to be selected. The final decision on
the combination of blood pumps and cannulae is to be reached by the
implanting surgeon, in consultation with Berlin Heart, Inc., Clinical
Affairs.
1000721x09 Revision 8
98
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...