Chapter 5 Description: Ikus
5.2.2 Display and operating panel
1
2
3
4
5
Fig. 5-3
Display and operating panel
1
Battery operation
The yellow LED lights up when the
Ikus
is running in battery operation
2 Alarm
The red LED lights up or flashes (design dependent) whenever there is an alarm message
which has not been acknowledged by the monitor program
3
Mute alarm (Audio paused
) Button to temporarily mute an alarm message. The message remains
displayed on the laptop.
4
Operating hour counter
The number of all operating hours is displayed. When the
Ikus
was in
operation for 2000 operating hours (maintenance interval) after the 1
st
commissioning respectively
after the last maintenance the display starts blinking and shows
S -0.0h
. The number of hours (i.e.
S -10.0h
) are displayed exceeding the maintenance interval. A maintenance sticker (on the right
side under the laptop holder) contains the information when maintenance is due.
5
Battery charge indicator
7 yellow LEDs indicate the battery charge state (see Fig. 5-3, page 61).
The red LED lights up when the batteries are completely discharged!
Fig. 5-4
Battery charge indicator and operating hour counter
The LEDs of the running time display correspond to the maximum technically possible
battery running time. However, the maximum permitted off-mains (battery) operating
time is 30 minutes. After 30 minutes of battery operation, a warning message
Batteries discharged - use power supply!
is displayed. Immediately switch the Ikus
over to mains operation. Failure to do so may result in failure of the device and
damage to the battery system.
5.2.3 Power supply
WARNING
The Ikus is designed for stationary operation and to be run on mains.
Do not run it on battery operation unless this is absolutely necessary
(e.g. when moving the patient within the clinic or during a mains failure).
To prevent the batteries from aging quickly, every period of battery
operation should be followed by at least 6 h of mains operation.
1000721x09 Revision 8
61
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...