Chapter 15 Troubleshooting and correcting faults
Connect new right blood pump to the Ikus
INSTRUCTION
1.
Fill the free ends of the cannulae with sterile saline solution. Make sure that all
air has been removed. Connect the prepared replacement pump to the cannu
lae.
2.
Plug the new driving tube into the freed connector. The plug snaps into place
clearly audible.
3.
Check that the plug is securely connected. To do so, grip the plug body above
the release sleeve and pull on it. Do not pull from the release sleeve, and
never from the tube!
4.
Release the tube clamps from the cannulae.
Starting the Ikus
INSTRUCTION
1.
Move the cursor to the field
step left
(RVAD) respectively
step right
(BVAD).
2.
RVAD: Confirm
Step left
with
<Enter>
to trigger a single step.
BVAD: Confirm
Step right
with
<Enter>
to trigger a single step.
3.
If any air bubbles are visible remove them via the de-airing needle. When all
air has been completely removed from the left pump: remove the de-airing
needle.
4.
Move cursor to the
OK
field and press
<Enter>
to confirm. The driving unit
starts up again using the defined parameters.
5.
Check whether the pump is filling correctly and, if necessary, adjust the
parameters.
6.
Secure all connections with cable ties. See section 9.11: Securing the connec
tions, page 130.
15.1.3 Replacing the left blood pump (LVAD/ BVAD)
WARNING
BVAD: If the left pump is being replaced, the right pump must also be
stopped while the pump is being replaced. Otherwise there is the risk
of pulmonary edema.
Material
•
1 prepared replacement blood pump (see section 15.1.1: Preparing a
replacement blood pump, page 195)
•
1 tube connecting set (cable tie, cable-tie gun), included in the accessory set.
Only the cable ties and cable tie guns provided should be used.
Stopping the left blood pump and detaching the blood pump from Ikus
INSTRUCTION
1.
Bring the patient into the Trendelenburg position.
2.
The cable tie covering the EXCOR cannula on the stub of the blood pump
should be removed carefully. Use an appropriate blunt tool. Important: never
use a sharp instrument, for example, a scalpel or scissors, to remove the
cable tie. This may cause damage to the cannula. Check cannulae immedi
ately to make sure they are not damaged.
1000721x09 Revision 8
197
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...