
Chapter 5 Description: Ikus
board runs autonomously and can be influenced neither by the 2 control computers
nor by the laptop.
In emergency pulse mode, the system will operate with the following settings:
Synchronous mode (biventricular)
Systol.
pressure
[mmHg]
Diastol.
pressure
[mmHg]
Rate
[bpm]
Relat. systole
duration [%]
left
210
-40
70
40
right
150
-40
70
40
Tab. 5-2
Settings in emergency pulse mode
5.5.4 Battery operation
If the mains fails, the batteries will supply power to the system for at least 30 minutes.
In this case, the battery operation indicator lights up. An acoustic alarm sounds at 10
minute intervals and a new message will be shown on the monitor program, showing
the time the system has already been running in battery operation. The LEDs on the
control panel of the handle show the charge state of the batteries.
5.5.5 Manual pump
If a working Ikus is no longer available, the blood pump(s) can also be temporarily
operated (even in BVAD mode) by the manual pump supplied with the system (see
section 15.5: Driving blood pump(s) with the manual pump, page 204).
5.5.6 Password-protected user profiles (access passwords)
Only a user who has logged into the monitor program with the correct password will
be allowed to change settings within the monitor program. When the system is
delivered, a single default user profile is configured; up to 9 additional user profiles can
be added. All system setting changes are logged specifically for the user who is
carrying them out.
1000721x09 Revision 8
65
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...