Chapter 15 Troubleshooting and correcting faults
3.
As soon as the original
Ikus
has stopped, remove the driving tube(s) from it. To
do so, take hold of the release sleeve and pull this out of the connector(s).
4.
Connect the driving tube(s) to the replacement
Ikus
.
IMPORTANT: Observe the colored markings. The sound of the plug snapping
into place is clearly audible. Check that the plug is securely connected. To do
so, grip the plug body above the release sleeve and pull on it. Do not pull from
the release sleeve, and never from the tube!
5.
To confirm the
OK
selection, press
<Enter>
. The replacement
Ikus
will start up
using the defined parameter settings.
6.
Check whether the pump is filling correctly and, if necessary, adjust the
parameters.
7.
Switch off the original Ikus (see section 6.3.3: Drive OFF: switching the Ikus
off, page 78).
15.5 Driving blood pump(s) with the manual pump
Fig. 15-1
Patient on manual pump
This is necessary if …
•
the power supply to the Ikus cannot be ensured
•
the Ikus has to be restarted (e.g. emergency operating mode) and there is no
replacement Ikus available
WARNING
The use of the manual pump is only permitted for medical personnel
trained in the use of it.
Pay attention to the colored markings on the driving tubes and on the
connectors of the manual pump. Otherwise, there is a risk of lung
edema.
Always keep manual pump attached to the Ikus. Otherwise in an
emergency situation the adequate support of the patient is not
guaranteed.
1000721x09 Revision 8
204
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...