Chapter 16 Appendix
When using staged cannulae or a cannula extension set / connecting set, the pumping
rate may not be greater than the respective value found, as the pump will not eject its
full volume at higher rates.
16.1.11Blood pump combinations in biventricular mode
left blood pump
right blood pump
10 ml
10 ml
15 ml
15 ml
30 ml
25 ml
60 ml
50 ml
Tab. 16-10
Recommended combinations
Check whether a blood pump combination that is not recommended has been
selected for the patient. The final decision on the combination of blood pumps and
cannulae is to be reached by the implanting surgeon, in consultation with Berlin Heart,
Inc Clinical Affairs.
16.1.12Relative systolic duration
The relative systolic duration is adjustable in the range of 20% and 70%. The upper
and lower threshold (20-30% and 60-70%) are marked in red on the laptop. For these
values it cannot be guaranteed that the activated pressure parameters are achievable
for each single case.
16.2 Technical specifications
Electro-pneumatic extracorporeal ventricular assist device
EXCOR
®
Pediatric VAD with Stationary Driving Unit Ikus
Manufactured by:
Berlin Heart GmbH
Wiesenweg 10
2247 Berlin
Germany
Classification
Class 3
Overall system (except sterile
products)
Ambient temperature in operation
+10 °C to +30 °C;
with restrictions of the battery performance
up to +35°C
Ambient temperature,
transportation and storage
-10 °C to +50 °C ;
6 h warming-up period before
commissioning after transportation
Tab. 16-11
Technical specifications
1000721x09 Revision 8
217
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...