Chapter 9 Implantation - surgical procedure
9.8
Shortening the cannulae if necessary
WARNING
If an EXCOR cannula extension set / connecting set is required for
implantation and the length of the tube part needs to be reduced, the
tube part should be cut but only to achieve the following minimum
lengths:
Article
Diameter /
Diameter reduction
Minimum length
Cannula extension set
A06-006
6 mm
55 mm
A09-009
9 mm
60 mm
A12-012
12 mm
75 mm
Connecting set
A06-009
9 to 6 mm
60 mm
A09-012
12 to 9 mm
75 mm
Tab. 9-2
Cannula extension set / connecting set: minimum length of tube section
INSTRUCTION
1.
Cut the cannulae to the required length. Make the cut perpendicular to the
cannula axis and ensure that the cut is straight.
2.
Make sure that the lengths of the 2 cannulae leading to the same pump match.
It must be possible to connect the cannulae to the pump without having to
exert any tension.
9.9
Connecting the blood pumps to the cannulae
WARNING
Ensure that cannulae, blood pump(s) and driving tubes are not subject
to external forces and are free of kinks or sharp bends.
When connecting the blood pump(s), pay attention to the direction of
the arrows on the inflow and outflow stubs. These show the direction of
the blood flow.
Type of support
Anastomosis of inflow
cannula to
Points upwards...
Univentricular
LVAD
apex
blood chamber
LVAD
atrium
air chamber
Tab. 9-3
Anastomosis and direction of the blood chambers
1000721x09 Revision 8
128
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...