T Lane Error Codes (Continued)
Table 4.6
Sensor
Problem
Solution
SL066A ON
SL066B ON
SL066 OFF
SL071 ON
SL071 OFF
SL072 ON
SL072 OFF
SL401 ON
SL401 OFF
SL402 ON
SL402 OFF
SL403 ON
SL403 OFF
SL404 ON
SL404 OFF
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that a carrier is not jammed or stuck.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
4. Select
Restart
on the control panel to resume
operation.
Troubleshooting
Error Code Tables
4
B31449AE
4-45
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...