Figure 1.1 Aliquot Module
1. Level Detector
2. Labeler
3. Aliquotter
Level Detector
The Level Detector measures the red cell layer or clot in the primary sample tube.
1. The system routes primary sample tubes to the Level Detector.
2. Grippers lower up to five tubes at one time into an optical well.
3. An optical sensor locates:
— the interface between the red cells and the silicon separator (gel) for gel tubes
Or
— the separation between the red cells and the serum or plasma for non-gel tubes.
4. The Level Detector measures the height from the bottom of the tube to the top of the
red cells.
5. The system stores this information and uses it later to calculate the volume of serum or
plasma available in the primary sample tube.
6. The grippers retrieve the primary sample tubes from the optical well and place them
back into sample tube carriers.
7. The track routes the primary sample tubes to the Labeler.
Power Express Aliquot Module Description
Components
1
B31449AE
1-3
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...