Use Bypass only to release empty sample carriers (except as otherwise noted in error
recovery). Samples do not route properly and cannot be reentered or remapped.
1
Select
Manual
and
OK
to place the module in Manual mode.
2
Select
Maintenance
on the control panel.
3
Select
Bypass
.
4
Wait for all sample tube carriers to pass through the module.
5
Select
Enter
and select
Exit
.
6
Select
Auto
, then select
OK
to resume processing.
Stop Button Recovery
When you press any yellow or white
Stop
button (except the Centrifuge
Stop
button), at any
time during operation, the system disconnects power to the Power Express system. The
Line Control and Cennexus remain
ON
. The air system also remains on.
Note
If you press the Centrifuge Stop button, refer to the Centrifuge Stop Button Recovery
Procedure, in the
Centrifuge Module IFU, PN B31456
.
Important
Press the Stop button to halt any hazardous condition.
Important
When you press the Stop button, the Line Control cannot accurately represent the
location of the last sample tube sorted by the system.
Caution
If you have an Aliquot module, when you press a Stop button, aliquot tips containing
sample drip very slowly. You may need to clean the drip trays. For more information,
refer to CHAPTER 5, Daily Maintenance Procedures, in the
General System Operation
IFU, PN B31448
.
Important
Perform only the Stop button recovery steps that apply to your Power Express
configuration.
Error Recovery Procedures
Power Express Aliquot Error Recovery Procedures
3-6
B31449AE
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...