background image

b. Perform this procedure from the Direct Track Sampling module control panel.

Select 

Maintenance

 

 on the control panel.

c. Select the number field to display the onscreen keyboard.
d. Enter function code 

91

 and select 

Enter

 

 twice to hold tubes at bar code reader

BR03 and to release tubes after bar code reader BR03. This ensures that the system
reads and registers the bar code IDs at BR03.

e. As the samples pass through the DTS sampling position make a note of the sample

ID numbers for later retrieval. See Note below.

f. After the system releases all tubes from BR03 and BR02, select

 Enter

 

 and 

Exit

 

.

g. Select 

Auto

 

 and 

OK

 

 to resume operation.

h. Perform troubleshooting on the connected analyzer as necessary to re-establish

communication with the Direct Track Sampling module.

Note

The analyzer does not aspirate samples exiting the DTS connection. If the

analyzer is a dynamic analyzer (AU680, DxC 700 AU, or DxI), Cennexus

automatically recognizes that the analyzer has not aspirated the samples and

reroutes them to another analyzer for testing. If the analyzer is a non-dynamic

analyzer, the system considers aspiration to be complete, and you must

manually request retesting or retrieve samples using Cennexus for offline

testing.

i. Use the Cennexus retrieval function to manually collect the samples, or use the LIS

console to order retests on the samples noted in step e above.

20

If the Abbott i2000SR Direct Track Sampling module T-Lane generates a CCIE error
when you press the main track Stop button:
a. Select 

Mute

 

 and 

Clear

 

 on the control panel to clear the errors.

b. Select 

Restart

 

 on the control panel to resume operation.

Note

Pressing the main track Stop button does not affect the Abbott i2000SR,

because the Abbott i2000SR is on a separate power system, but doing so does

generate the CCIE error.

21

At the T-Lanes, if a stopper is jamming the sample tube carrier, push it forward to
release the carrier. Leave other tubes and sample tube carriers alone.

22

At the Aliquotter and Labeler components:

Note

Regardless of the status of the Aliquotter and Labeler, the recovery procedure is

the same.

Error Recovery Procedures

Power Express Aliquot Error Recovery Procedures

3

B31449AE

3-13

Summary of Contents for Power Express Aliquot

Page 1: ...Instructions for Use Power Express Aliquot Module For In Vitro Diagnostic Use B31449AE May 2018 Beckman Coulter Inc 250 S Kraemer Blvd Brea CA 92821 U S A...

Page 2: ...marks mentioned herein are trademarks or registered trademarks of Beckman Coulter Inc in the United States and other countries All other trademarks are the property of their respective owners Find us...

Page 3: ...roughout the IFU Updated the Power Express Aliquot Module Key Features section Chapter 2 Aliquot Module Procedures Updated the Set Load Timeout for the Level Detector section Updated the Replace the R...

Page 4: ...e Function Codes Function Code Procedure General Error Recovery Stop Button Recovery Aliquot Drip Tray Error Recovery Updated Error and Function codes Updated Added diagrams B31449AC 06 2015 Cennexus...

Page 5: ...ulter replaces any part under this warranty or as a result of any Services performed Beckman Coulter will own the replaced part Your exclusive warranty for Services not performed in a workmanlike mann...

Page 6: ...Warranty and Service vi B31449AE...

Page 7: ...ntially hazardous situation which if not avoided could cause death or serious injury Warning can indicate the possibility of erroneous data that could cause an incorrect diagnosis Caution Caution indi...

Page 8: ...te the system only with all covers in place Caution Do not load or view the Instructions for Use PDF files onto any computer connected to the automation system Failure to follow this caution can reduc...

Page 9: ...y or biohazard Please Hold This Portion Label This label located on the drawer in the Dynamic Inlet and Outlet modules and the top cover of the centrifuge indicates the most stable position to hold th...

Page 10: ...s and materials declaration table the Table of Hazardous Substance s Name and Concentration meet People s Republic of China Electronic Industry Standard SJ T11364 2006 Marking for Control of Pollution...

Page 11: ...n assessed before being placed on the market and has been found to meet the applicable directives relating to the European Union safety health and environmental protection requirements RoHS Caution Sy...

Page 12: ...Symbol Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot for a variety of reasons be presented on t...

Page 13: ...ds and lab coats Handle and dispose of biohazardous materials according to your laboratory procedures Title of Symbol Warning Crushing of hands Meaning of Symbol To warn of a closing motion of mechani...

Page 14: ...l device labels labelling and information to be supplied Part 1 General Requirements 5 1 2 Title of Symbol Manufacturer Meaning of Symbol Indicates the medical device manufacturer as defined in EU Dir...

Page 15: ...nd off button for the device Title of Symbol OFF power Meaning of Symbol To indicate disconnection from the mains at least for mains switches or their positions and all those cases where safety is inv...

Page 16: ...labelling and information to be supplied Part 1 General Requirements 5 1 7 Title of Symbol Stop Meaning of Symbol To identify the control or the indicator to stop the active function Standard Number T...

Page 17: ...ental electronic information products This logo indicates that this electronic information product does not contain any toxic or hazardous substances or elements and is green and is environmental The...

Page 18: ...r crush This label is found in several locations Caution parts move automatically While the system is in operation do not touch or go close to any moving parts Close protective guards and covers durin...

Page 19: ...econdary Tube Container 2 6 Add Aliquot Tips 2 8 Empty Aliquot Tip Biohazard Container 2 9 CHAPTER 3 Error Recovery Procedures 3 1 Introduction 3 1 Error Recovery General Information 3 2 Aliquot Modul...

Page 20: ...otter Error Codes 4 21 Aliquot Merge Lane 4 38 T Lane Error Codes 4 39 Sensor Diagrams 4 46 Level Detector Unit 4 48 Labeler Unit 4 49 Aliquot Unit 4 52 APPENDIX A Related Documents A 1 Related Docume...

Page 21: ...wer Express Aliquot module This manual contains information and instructions to perform Power Express Aliquot module operations and troubleshooting functions Note Follow all safety cautions and warnin...

Page 22: ...Introduction Scope of this Manual xxii B31449AE...

Page 23: ...an LIS assigned sample ID Automatically prints and applies aliquot tube bar code labels decreasing the possibility of incorrect labels Labels can be printed according to content as described in CLSI A...

Page 24: ...quot module includes three main hardware components Level Detector Labeler Aliquotter The three components of the Aliquot module work in sequence to process the sample tubes 1 The Level Detector measu...

Page 25: ...tor gel for gel tubes Or the separation between the red cells and the serum or plasma for non gel tubes 4 The Level Detector measures the height from the bottom of the tube to the top of the red cells...

Page 26: ...further information please contact your Beckman Coulter representative Important The secondary sample tubes and tube caps supplied with the automation system are not Department of Transportation DOT o...

Page 27: ...Labeler checks the bar code on each label 6 If the Labeler cannot read the bar code a It discards the secondary tube into a chute for disposal in the defective secondary tube container b It repeats St...

Page 28: ...econdary tubes the Aliquotter uses one of the following user defined priority settings to 1 create as many secondary tubes as possible regardless of aliquot priority 2 create as many secondary tubes a...

Page 29: ...iquot tip aspirates the total serum volume required to fill all the secondary sample tubes 3 The Aliquotter reads the bar code of each secondary tube and verifies it to be associated with the primary...

Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...

Page 31: ...the Level Detector The Level Detector loads and measures up to five primary sample tubes at a time The Load Timeout parameter is the period of time in seconds that the Level Detector waits before it...

Page 32: ...low Figure 2 1 Aliquot Module Labeler 1 Aliquot Sample Tube Supply Bin 2 Label Assembly Drawer 2 Confirm that the label roll has a sufficient supply of labels for the run To replace a roll of labels r...

Page 33: ...e back 3 Remove the empty spool or feed through any paper that is left on the label roll 4 Check the inside of the printer for dust If necessary clean the inside of the printer and the print head a Re...

Page 34: ...ines at the bottom For ease of installation remove seven labels from the start of the roll 7 Insert the start of the new roll of labels through the label paper intake slot at the front of the printer...

Page 35: ...of the U clamp Push the U clamp into the slots on the uptake spindle 9 Take up any paper slack by manually winding the label paper clockwise on the uptake spindle 10 Close the printer door and press t...

Page 36: ...the bottom of the line inside the bin Do not fill to capacity 3 Close the door 4 Select the warning message on the control panel to clear the warning and then resume operation Empty the Defective Seco...

Page 37: ...amaged aliquot tubes c Place undamaged unlabeled aliquot tubes back into the aliquot tube supply bin Refer to Figure 2 5 Aliquot Module Labeler Figure 2 5 Aliquot Module Labeler 1 Label Assembly Drawe...

Page 38: ...message 1 Select the warning message on the control panel to silence the alarm 2 Open the aliquot tip supply bin door and check the supply level Refer to Figure 2 6 Aliquotter Consumables Figure 2 6...

Page 39: ...zard procedures 1 Select the warning message on the control panel to silence the alarm 2 Open the aliquot tip disposal door at the lower front of the Aliquotter Refer to Figure 2 7 Aliquot Tip Biohaza...

Page 40: ...he door Note If you do not push the drawer all the way back in the system displays a No Pail Error on the control panel Check for loose tips or other obstructions behind the drawer 7 Select Restart to...

Page 41: ...laboratory policies and procedures related to the handling of biohazardous materials Refer to the applicable sources such as Material Safety Data Sheets for specific hazard information Caution Before...

Page 42: ...tion on a connected instrument The following table lists the Power Express module function codes by associated module with descriptions of the action triggered by each function code You must place a m...

Page 43: ...ance to disable drip trays Function Code Procedure 1 Select Manual and OK to place the module in Manual mode 2 Select Maintenance on the control panel 3 Select the number field to display the onscreen...

Page 44: ...ecovery Procedures This section provides information to identify and resolve errors related to system operational problems that are not specific to any Power Express module System Error Recovery proce...

Page 45: ...r Contact your Beckman Coulter representative for assistance Jammed Sample Tube Carrier Error Recovery To recover from a jammed sample tube carrier error follow this procedure 1 Select Mute on the con...

Page 46: ...ain ON The air system also remains on Note If you press the Centrifuge Stop button refer to the Centrifuge Stop Button Recovery Procedure in the Centrifuge Module IFU PN B31456 Important Press the Sto...

Page 47: ...d RF02 positions e Determine which rack or Bulk Inlet the samples removed in steps c and d were loaded from Reload the sample tubes to racks in the Inlet module or the Bulk Inlet bin 5 At the Error La...

Page 48: ...ld Repeat and manually move both arms to an accessible location Remove sample tubes from Loading and Retrieval Arm grippers Use a flexible claw like grabber tool to reach inside the module to retrieve...

Page 49: ...the diagrams below Note Any samples removed from a Storage module with any floor configured for Sarstedt tubes the sample tube cannot be reloaded You must remove the sample tube from the track and sto...

Page 50: ...N033 FS002 ON FS004 ON FS006 ON FS032 ON FS034 ON or FS036 ON error Remove the tubes from the sample tube carrier at SN003 or SN033 When the module is placed back into Auto mode the empty sample tube...

Page 51: ...es in the Unload rack so that the tubes start at position 1 and have no empty spaces e Block any tubes behind RF01 position Select Auto and OK to place the module back into operation Allow the system...

Page 52: ...this rack s in the output tray on the right side of the instrument s d Remove sample tubes from the instrument load shuttle Place these tubes into another rack and place this rack in the output tray o...

Page 53: ...reroutes them to another analyzer for testing If the analyzer is a non dynamic analyzer the system considers aspiration to be complete and you must manually request retesting or retrieve samples using...

Page 54: ...layed dispose of the aliquot tip down the disposal chute 4 Unlock the rear doors of the Aliquotter and open the rear doors 5 Collect the sample by placing a 2ml sample cup below aliquot tip 1 and then...

Page 55: ...PS043 is generated for the sample tube that was at BR01 Confirm the sample ID noted in step a manually remove the cap e Select Clear from the control panel to clear the PS041 PS042 or PS043 and relea...

Page 56: ...push it forward to release the carrier c Press the Stop button again This resets the Stop button d Turn the Centrifuge power switch OFF and wait at least a minute The system displays Unlock when you...

Page 57: ...procedure on the Centrifuge D Lane control panel Select Maintenance on the control panel r Select the number field to display the onscreen keyboard s Enter function code 84 and select Enter twice t A...

Page 58: ...e Aliquot component b Select the number field to display the onscreen keyboard c Enter code 32 and select Enter twice on the control panel d If TIP ERR 12 TIP ERR 13 or TIP ERR 23 occurs select Clear...

Page 59: ...to hold tubes at bar code reader BR02 and to release tubes after BR02 f Remove all tubes in the Aliquotter and the Labeler Do not remove any sample tubes at BR02 or before the tube is in the Labeler...

Page 60: ...ld to display the onscreen keyboard c Enter code 32 and select Enter twice on the control panel d If TIP ERR 12 TIP ERR 13 or TIP ERR 23 occurs select Clear and OK on the control panel If the Clear op...

Page 61: ...the cause of the error The Solution column provides resolution instructions Nomenclature The following table contains definitions of the alphabetic portion of error codes found in the Error Code colu...

Page 62: ...X Printer Syntax PRNT CUT Printer Cut PRNT HEAT Printer Heat PRNT PAUSE Printer Pause PS Photo Sensor REG RST Register Reset RF RFID RF INI RF Initialization RST Reset SL Cylinder SN Sensor TIP ERR Ti...

Page 63: ...the carrier CCIE ERR Established communication between PLC and LC has been disrupted 1 Confirm the cable connection between LC and PLC located at the upper stream side of the Error unit 2 Confirm all...

Page 64: ...ng arm 3 Place the module in Auto mode to resume operation Note If the error still occurs reboot the module then place it in Auto mode to resume operation PM001 LIM The current register setting forced...

Page 65: ...Select Restart on the control panel to resume operation SL011A ON SL011B ON SL011 OFF Error at the Y axis cylinder on the carrier transfer arm ON error The cylinder does not respond correctly to an ON...

Page 66: ...ch is working correctly 2 Confirm that the compressed air supply is within the normal pressure range SL093 ON SL093 OFF Locking cylinder error The cylinder that locks the front door lock is not workin...

Page 67: ...to resume operation SN002 PAS SN003 PAS SN004 PAS SN005 PAS SN102 PAS SN202 PAS SN204 PAS SN206 PAS Passage sensor error A passage sensor does not detect a carrier within a predetermined timeout peri...

Page 68: ...of an aliquot tube either here or somewhere else The sample tube will need to be removed and select Clear to recover from the error 03 indicates the sample tube bar code label was read however the ID...

Page 69: ...moving to the home position 1 Place the module in Manual mode 2 Confirm that nothing interferes with the movement of the rotating table 3 Place the module in Auto mode to resume operation PM001 JOG H...

Page 70: ...he Y axis inner arm or retrieve arm motor moves towards the home position and the home sensor works correctly 2 Confirm that nothing interferes with the movement of the loading arm 3 Place the module...

Page 71: ...ss Label Feed button on the Label Printer Refer to figure 2 3 and rotate spindle to take up any loose label paper Close drawer and press Restart PRNT STATE Failed to send data to the printer 1 Confirm...

Page 72: ...sor at the carrier does not detect the tube 1 Confirm that PS024 works correctly 2 If a secondary container was dropped pick up the tube and place it in the carrier at PS024 PS031 ON Jammed or stuck s...

Page 73: ...S051 works correctly 2 If there is a secondary container at PS051 remove the tube from the carrier PS051 OFF Secondary container detection error at BR03 1 Confirm that PS051 works correctly 2 Tube was...

Page 74: ...secondary container is level and is not jammed or stuck 2 Confirm that the air compressor is on and the auto switch is working correctly 3 Confirm that the compressed air supply is within the normal...

Page 75: ...or is on and the auto switch is working correctly 3 Confirm that the compressed air supply is within the normal pressure range SL006 ON SL006 OFF Secondary container guiding clamps are not working cor...

Page 76: ...al pressure range SL021 ON SL021 OFF Error at the gripper on the secondary container transfer arm ON error The cylinder does not respond correctly to an ON signal from a PLC OFF error The cylinder doe...

Page 77: ...o switch is working correctly 3 Confirm that the compressed air supply is within the normal pressure range SL034 ON SL034 OFF Push up cylinder error inside the secondary container supply box ON error...

Page 78: ...ctions 2 Confirm that the air compressor is on and the auto switch is working correctly 3 Confirm that the compressed air supply is within the normal pressure range SL041 ON SL041 OFF SL042 ON SL042 O...

Page 79: ...air compressor is on and the auto switch is working correctly SL086A ON SL086B ON SL086 OFF Error at the Z axis cylinder on the carrier transfer arm ON error The cylinder does not respond correctly t...

Page 80: ...Confirm that SN011 turns on when the drawer is secured in position 3 If error repeats press the Label Feed button on the Label Printer Refer to figure 2 3 and rotate spindle to take up any loose label...

Page 81: ...If necessary reprint the bar code label and re apply it to the sample tube 3 If the error persists select Clear on the control panel to release the carrier The aliquot process and all sample tubes in...

Page 82: ...It processed Tube with aliquoting error HI VOL2 Unexpected serum level was detected Logged as an aliquoting error in System Error Log HI VOL2_S A sensor detected an impact when a nozzle attempted to...

Page 83: ...on Note If the error still occurs reboot the module then place it in Auto mode to resume operation PM001 POS Loading arm positioning error 1 Place the module in Manual mode 2 Confirm that nothing inte...

Page 84: ...hing interferes with the movement of the loading arm 3 Place the module in Auto mode to resume operation PM002 HOM Error generated while the arm was moving to the home position 1 Place the module in M...

Page 85: ...ated while the arm was moving to the home position 1 Place the module in Manual mode 2 Confirm that nothing interferes with the movement of the loading arm 3 Place the module in Auto mode to resume op...

Page 86: ...module in Manual mode 2 Confirm that nothing interferes with the movement of the loading arm 3 Select Restart on the control panel to resume operation PM101 ERR Motor error 1 Place the module in Manu...

Page 87: ...hat nothing interferes with the movement of the loading arm 3 Select Restart on the control panel to resume operation PM102 LIM Motor error 1 Place the module in Manual mode 2 Confirm that nothing int...

Page 88: ...Select Restart on the control panel to resume operation PS004 OFF Tip detection error at Nozzle A The sensor did not detect a tip on the nozzle when arm A attempted to discard it down the duct 1 Confi...

Page 89: ...k of the tip supply box Inspect for and remove jammed or stuck tips PS086 ON Jammed or stuck tip at the front of the tip supply box Inspect for and remove jammed or stuck tips PS104 OFF Tip detection...

Page 90: ...within the normal pressure range 4 Select Restart on the control panel to resume operation SL016A ON SL016B ON Error at the splash guard cover at the aspiration point 1 Inspect for and remove any obst...

Page 91: ...e range SL051 ON SL051 OFF Error at lateral tube stopper 1 ON error The cylinder does not respond correctly to an ON signal from a PLC OFF error The cylinder does not respond correctly to an OFF signa...

Page 92: ...or The cylinder does not respond correctly to an ON signal from a PLC OFF error The cylinder does not respond correctly to an OFF signal from a PLC 1 Confirm that a carrier is not jammed or stuck and...

Page 93: ...rror The cylinder does not respond correctly to an ON signal from a PLC OFF error The cylinder does not respond correctly to an OFF signal from a PLC 1 Confirm that a carrier is not jammed or stuck 2...

Page 94: ...essor is on and the auto switch is working correctly 3 Confirm that the compressed air supply is within the normal pressure range 4 Select Restart on the control panel to resume operation SL091 ON SL0...

Page 95: ...l from a PLC OFF error The cylinder does not respond correctly to an OFF signal from a PLC 1 Confirm that a carrier is not jammed or stuck 2 Confirm that the air compressor is on and the auto switch i...

Page 96: ...uard cover at the aspiration point 1 Inspect for and remove any obstructions 2 Confirm that the air compressor is on and the auto switch is working correctly 3 Confirm that the compressed air supply i...

Page 97: ...TIPFALL NG B Discarded tip not detected in tip disposal bin sensor PS058 1 Select the warning message on the control panel to silence the alarm 2 Inspect and remove any obstructions from the tip disp...

Page 98: ...auto switch is working correctly 3 Confirm that the compressed air supply is within the normal pressure range 4 Select Restart on the control panel to resume operation SL106A ON SL106B ON SL106 OFF St...

Page 99: ...XX is unable to determine how to handle the sample tube and why it is unable to handle the sample tube 0y 01 indicates the sample tube bar code label is unreadable Repeat the read operation or reprin...

Page 100: ...estart on the control panel to resume operation QJ71GF ERR The PLC module QJ71GF used for the communication between the PLC and LC malfunctioned 1 Reset the PLC 2 If the reset does not recover from th...

Page 101: ...m that the compressed air supply is within the normal pressure range 4 Select Restart on the control panel to resume operation SL013 ON SL013 OFF Diverter error The diverter is not working correctly O...

Page 102: ...in the normal pressure range 4 Select Restart on the control panel to resume operation SL025A ON SL025B ON SL025 OFF Diverter error The diverter is not working correctly ON error The cylinder does not...

Page 103: ...d air supply is within the normal pressure range 4 Select Restart on the control panel to resume operation SL045A ON SL045B ON SL045 OFF Diverter error The diverter is not working correctly ON error T...

Page 104: ...does not respond correctly to an ON signal from a PLC OFF error The cylinder does not respond correctly to an OFF signal from a PLC 1 Confirm that a carrier is not jammed or stuck 2 Confirm that the a...

Page 105: ...ror The cylinder does not respond correctly to an ON signal from a PLC OFF error The cylinder does not respond correctly to an OFF signal from a PLC 1 Confirm that a carrier is not jammed or stuck 2 C...

Page 106: ...at a carrier is not jammed or stuck and that the belt moves 2 Select Restart on the control panel to resume operation 3 If no sample tube carrier is there or stopped the passage of the carrier did not...

Page 107: ...has occurred and then look up the error code in the Error Code tables in this document The most common errors tend to be jammed sample tube carriers which can happen anywhere along the track Sample tu...

Page 108: ...4 1 Level Detector Sensor Diagram SN031 AS021 SL021 SN011 SN013 PM001 SL014 AS022 SN032 SN033 SN034 SN035 2 3 5 4 1 AS019 AS017 AS015 AS016 AS018 AS012 SL011B AS011 SL013 AS014 SL012 AS013 SL011A 4 4...

Page 109: ...04 SN062 BR03 BR04 SN066 SN063 SL043 AS043 c SL042 AS042 AS041 SL041 b a SL044 DM041 DM041 SN064 SN054 SL053 AS053 f SL052 AS052 AS051 SL051 e d DM051 DM051 FS041 SN043 PS061 RF03 SN053 i g h SN061 SN...

Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...

Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...

Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...

Page 113: ...AS021 SL021 B SL121 AS121 d A SL022 AS022 AS122 SL122 B SL025 A SL125 B AS123 SL123 SL124 DM021 A DM121 B AS023 SL023 f B AS126 SL126A A AS026 SL026A SL024 B A A B SL126B SL026B h B AS131 SL131 A SL03...

Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...

Page 115: ...001 AS002 SL003B SL003A SL103B SL103A SN001 SN002 Z2 PM103 PM003 PM002 PM102 SN101 SN102 Z1 A B A B PRS11 AS104 AS105 SL104A SL104B PS104 PS004 PRS01 SN103 SN003 AS004 AS005 SL004A SL004B SN104 SN004...

Page 116: ...43 SL043 l k AS004 SL004 AS003 SL003 x w AS042 SL042 AS041 SL041 5 4 3 2 1 SN062 y z SN061 A A Y SN066 SN067 SL066a SL066b AS064A SL064a AS064B SL064b SL067a SL067b AS067 AS068 PS063 PS061 PS065 PS062...

Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...

Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...

Page 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...

Page 120: ...Related Documents Related Documents A 2 B31449AE...

Page 121: ......

Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...

Reviews: