Level Detector Error Codes (Continued)
Table 4.2
Sensor
Problem
Solution
AS015 OFF
AS016 OFF
AS017 OFF
AS018 OFF
AS019 OFF
Gripper error on the
transfer arm.
1. Confirm that a sample tube has not been dropped.
2. Confirm that the gripper is holding a sample tube correctly.
If not, manually place a tube in the gripper. If due to a tube
removed for a BUZZ error at BR01, remove the tube after
the level sense operation is completed.
3. Confirm that the air compressor is on and the auto switch
is working correctly.
4. Confirm that the compressed air supply is within the
normal pressure range.
BUZZ01-0*
Line Control CPU Error, bar
code label, or RFID read
error or Sample Positioning
error.
1. Select
Restart
on the control panel to retry.
2. Confirm that the bar code reader or RFID reader reads the
bar code label or RFID tag correctly. If necessary, reprint
the bar code label and re-apply it to the sample tube.
3. If the error persists, select
Clear
on the control panel
to release the carrier.
CCIE ERR
Established communication
between PLC and LC has
been disrupted.
1. Confirm the cable connection between LC and PLC, located
at the upper stream side of the Error unit.
2. Confirm all the PLCs within the line are powered on.
3. When the communication between PLC and LC has been
re-established, an error is recovered.
4. Select
Restart
on the control panel to retry or use
Function code 01 (Lane through (Bypass)) to continue.
CCIE SND
Failed to send a message to
the Line Control.
Confirm that the cables between individual PLCs and the Line
Controller are connected correctly.
NETC1 DATA
Network protocol
converter: command
execution error.
Clear the error to retry the command. If the error persists,
reboot the module.
NETC1 RST1
NETC1RST2
Network protocol
converter: reminder to
reboot the module. After
replacing a network
protocol converter and
configuring its settings for
the first time, the module
must be rebooted for the
settings to take effect.
Reboot the module.
OP CALL
PLC hardware error. A PLC
battery may be worn out.
Replace the battery if the battery indicator on the front of the
PLC box is blinking.
Troubleshooting
Error Code Tables
4
B31449AE
4-3
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...