CHAPTER 2
Power Express Aliquot Module Procedures
Introduction
This chapter contains operational procedures for the Aliquot module.
Warning
Always operate the system with all shields and doors in position and secured to avoid
injury.
Perform system operations with caution.
Wear Personal Protective Equipment (PPE) such as gloves, eye shields, and lab coats.
Wash hands thoroughly after contact with sample media and all maintenance
activities.
Observe all laboratory policies and procedures related to the handling of
biohazardous materials.
Refer to the applicable sources (such as Material Safety Data Sheets) for specific
hazard information.
Set Load Timeout for the Level Detector
The Level Detector loads and measures up to five primary sample tubes at a time. The Load
Timeout parameter is the period of time (in seconds) that the Level Detector waits before it
processes less than a full load of five primary sample tubes.
Follow the procedure below to set the Load Timeout.
1
Select
Manual
on the control panel.
2
Select
OK
to stop the Level Detector.
3
Select
Maintenance
on the control panel.
4
Select the number field to display the onscreen keyboard.
5
Enter
82
from the onscreen keyboard and select
Enter
twice to enter the edit mode.
6
Enter the new setting in 10 second increments (for example, enter
6
for 60 seconds)
and select
Enter
twice to return to the Mode-Setting screen. Default setting is 6 for
60 seconds.
7
Select
Exit
to display the Status/Mode screen.
8
Select
Auto
and
OK
to resume operation.
B31449AE
2-1
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...