Labeler Error Codes (Continued)
Table 4.3
Sensor
Problem
Solution
BUZZ01-0*
BUZZ02-0*
BUZZ03-0*
BUZZ04-0*
Line Control CPU Error, bar
code label, or RFID read
error or Sample Positioning
error.
1. Select
Restart
on the control panel to retry.
2. Confirm that the bar code reader or RFID reader reads the
bar code label or RFID tag correctly. If necessary, reprint
the bar code label and re-apply it to the sample tube.
3. If the error persists, select
Clear
on the control panel
to release the carrier.
-01 indicates the sample tube bar code label is unreadable.
Repeat the read operation or reprint the bar code label and re-
apply it to the tube and repeat the read operation.
-02 indicates the sample tube bar code label was read, however
the ID indicates this tube should be somewhere else on the
system. This is frequently due to a mis-read of an aliquot tube
either here or somewhere else. The sample tube will need to be
removed and select Clear
to recover from the error.
-03 indicates the sample tube bar code label was read, however
the ID indicates the tube should not be on the system at all.
This is frequently due to a secondary label which may be
partially in view. Correct any secondary label and repeat, and if
the problem continues, remove the sample tube and select
Clear
.
FS031 ON
Stuck secondary container
at the tube delivery belt.
1. Inspect for a stuck secondary container on the tube
delivery belt.
2. If found, push the tube towards the container transfer
point.
ID ERR
Invalid secondary container
ID.
1. Remove all sample tubes from the Aliquot and Labeler
modules using the procedure
Aliquotter and Labeler Lane
Clear
.
2. Place the module in
Auto
mode to resume operation.
NETC1 DATA
Network protocol
converter: command
execution error.
Clear the error to retry the command. If the error persists,
reboot the module.
Troubleshooting
Error Code Tables
4-8
B31449AE
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...