Labeler Error Codes (Continued)
Table 4.3
Sensor
Problem
Solution
PS081 PAS
PS082 PAS
PS083 PAS
Passage sensor error. A
passage sensor does not
detect a carrier within a
predetermined timeout
period.
1. Confirm that a carrier is not jammed or stuck and that the
belt moves.
2. Select
Restart
on the control panel to resume
operation.
3. If no sample tube carrier is there or stopped, the passage
of the carrier did not trigger the sensor. Use a metal object
to trigger the sensor manually. Select
Restart
.
RF03 INI
RF04 INI
RFID initialization error.
1. Confirm that the RFID cable is connected correctly.
2. Reboot the module.
3. Place the module in
Run
mode to resume operation.
SL001 ON
SL001 OFF
Secondary container clamps
at the labeling position are
not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that the secondary container is level and is not
jammed or stuck.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
SL002 ON
SL002 OFF
Cylinder error at tube
pusher #1 that pushes out a
tube.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Inspect for and remove any obstructions.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
SL003 ON
SL003 OFF
Cylinder error at tube
pusher #2 that stops the
movement of a tube.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Inspect for and remove any obstructions.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
Troubleshooting
Error Code Tables
4-14
B31449AE
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......