Add Aliquot Tips
The Aliquotter transfers a portion of the sample from the primary sample tube to the
secondary sample tubes. If the component does not detect an aliquot tip, the system sounds
an alarm and activates a flashing beacon. The control panel turns yellow and a "Tip Supply"
message.
1
Select the warning message on the control panel to silence the alarm.
2
Open the aliquot tip supply bin door and check the supply level. Refer to
Figure 2.6
Aliquotter Consumables
.
Figure 2.6 Aliquotter Consumables
1. Aliquot Tip Supply Bin
2. Aliquot Tip Disposal Bin
3
Fill the bin to the bottom of the line inside the bin. Do not fill to capacity.
4
Close the door.
5
Select
Restart
to resume operation.
Power Express Aliquot Module Procedures
Add Aliquot Tips
2-8
B31449AE
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...