Labeler Error Codes (Continued)
Table 4.3
Sensor
Problem
Solution
PS034 ERR
Secondary container not
consistently detected.
1. Inspect for a stuck secondary container on the tube
delivery belt or obstruction around the tube transfer point.
2. If found, push the tube towards the tube transfer point or
remove obstruction.
3. Check for too many secondary containers on the tube
delivery belt. Should be no more than one past FS031. If
found, remove by pushing the tube upwards and place
back into the tube supply bin.
4. If the error still occurs, contact your Beckman Coulter
representative.
PS051 ON
Tube-on-tube error at
BR03. The transfer arm
attempted to place a
secondary container into
the carrier, but another
secondary container is in
the carrier.
1. Confirm that PS051 works correctly.
2. If there is a secondary container at PS051, remove the tube
from the carrier.
PS051 OFF
Secondary container
detection error at BR03.
1. Confirm that PS051 works correctly.
2. Tube was dropped. Place secondary tube into carrier at
PS051.
PS061 ON
Tube-on-tube error at
BR04. The transfer arm
attempted to place a
secondary container into
the carrier, but another
secondary container is in
the carrier.
1. Confirm that PS061 works correctly.
2. If there is a secondary container at PS061, remove the tube
from the carrier.
PS061 OFF
Secondary container
detection error at BR04.
1. Confirm that PS061 works correctly.
2. Tube was dropped. Place secondary tube into carrier at
PS061.
PS081 ON
PS082 ON
PS083 ON
A carrier is under the
transfer arm.
1. Inspect to see if there is a carrier under the carrier transfer
arm. If found, remove the carrier from the area.
2. Confirm that the sensor is working correctly.
Troubleshooting
Error Code Tables
4
B31449AE
4-13
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...