If necessary, select the warning message on the control panel to silence the alarm.
3
Check the tubes in the defective secondary tube container and dispose as follows:
a. Discard any labeled aliquot tubes.
b. Discard any damaged aliquot tubes.
c. Place undamaged, unlabeled aliquot tubes back into the aliquot tube supply bin.
Refer to
Figure 2.5 Aliquot Module Labeler
.
Figure 2.5 Aliquot Module Labeler
1. Label Assembly Drawer
2. Defective Secondary Tube Container
4
Replace the container and close the door.
5
Select
Restart
to resume operation.
Power Express Aliquot Module Procedures
Empty the Defective Secondary Tube Container
2
B31449AE
2-7