Aliquotter Error Codes (Continued)
Table 4.4
Sensor
Problem
Solution
SL103A ON
SL103B ON
Nozzle picker error.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that the air compressor is on and the auto switch
is working correctly.
2. Confirm that the compressed air supply is within the
normal pressure range.
SL104A ON
SL104B ON
SL104 OFF
The cylinder used to extend
the drip tray under aliquot
arm A has an error.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that the air compressor is on and the auto switch
is working correctly.
2. Confirm that the compressed air supply is within the
normal pressure range.
SL111 ON
SL111 OFF
SL112 ON
SL112 OFF
SL113 ON
SL113 OFF
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that a carrier is not jammed or stuck.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
4. Select
Restart
on the control panel to resume
operation.
SL116A ON
SL116B ON
Error at the splash guard
cover at the aspiration
point.
1. Inspect for and remove any obstructions.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
SL121 ON
SL121 OFF
SL122 ON
SL122 OFF
SL123 ON
SL123 OFF
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that a carrier is not jammed or stuck.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
4. Select
Restart
on the control panel to resume
operation.
Troubleshooting
Error Code Tables
4
B31449AE
4-35
Summary of Contents for Power Express Aliquot
Page 6: ...Warranty and Service vi B31449AE...
Page 22: ...Introduction Scope of this Manual xxii B31449AE...
Page 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Page 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Page 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Page 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Page 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Page 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Page 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Page 120: ...Related Documents Related Documents A 2 B31449AE...
Page 121: ......
Page 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...