A.S.T. Angewandte SYSTEM-TECHNIK GmbH Dresden
Page
27
b A S I
C
3.1.1.11. Stable number “StAN”
Set in terms of numbers in the range
between values from 10 … 250.
This value defines the number of
averaged ADC values taken into account
for testing the stable condition.
A higher number provides a safer stable
condition but it also extends the minimum
time necessary to recognize the stable
condition after a load has been changed.
default setting: 50
3.1.1.12. Stable range “StAR”
Set in terms of tenth of a division in the
range of 1 to 255.
This value defines the range of tolerance
a weighing sample has match in order to
meet the stable condition.
A higher value provides a safer and
faster stable condition.
default setting: 10
3.1.1.13. Limit underload “LUNL”
Set in terms of divisions in the range of
0 to 1000.
This value defines the number of divisions
below zero that have to be indicated
before the underload message has been
generated.
default setting: 9
3.1.1.14. Limit overload “LOVL”
Set in terms of divisions in the range of
0 to 1000.
This value defines the number of divisions
above FS (max.) that have to be indicated
before the overload message has been
generated.
default setting: 9
-1
+1
cursor
left
S t A N
0
/NET
GROSS
ENT
0 5 0
/NET
GROSS
0
SHIFT
>
FUNC
value
incr. decr.
cursor
right
TARE
-1
+1
cursor
left
S t A R
0
/NET
GROSS
ENT
0 1 0
/NET
GROSS
0
SHIFT
>
FUNC
value
incr. decr.
cursor
right
TARE
-1
+1
cursor
left
L U N L
0
/NET
GROSS
ENT
0 0 0 0 9
/NET
GROSS
0
SHIFT
>
FUNC
value
incr. decr.
cursor
right
TARE
TARE
-1
+1
cursor
left
L O V L
0
/NET
GROSS
ENT
0 0 0 0 9
/NET
GROSS
0
SHIFT
>
FUNC
value
incr. decr.
cursor
right
Summary of Contents for a810
Page 2: ......